Phase contrast magnetic resonance imaging explained

Phase contrast magnetic resonance imaging
Purpose:method of magnetic resonance angiograph

Phase contrast magnetic resonance imaging (PC-MRI) is a specific type of magnetic resonance imaging used primarily to determine flow velocities. PC-MRI can be considered a method of Magnetic Resonance Velocimetry. It also provides a method of magnetic resonance angiography. Since modern PC-MRI is typically time-resolved, it provides a means of 4D imaging (three spatial dimensions plus time).[1]

How it Works

Atoms with an odd number of protons or neutrons have a randomly aligned angular spin momentum. When placed in a strong magnetic field, some of these spins align with the axis of the external field, which causes a net 'longitudinal' magnetization. These spins precess about the axis of the external field at a frequency proportional to the strength of that field. Then, energy is added to the system through a Radio frequency (RF) pulse to 'excite' the spins, changing the axis that the spins precess about. These spins can then be observed by receiver coils (Radiofrequency coils) using Faraday's law of induction. Different tissues respond to the added energy in different ways, and imaging parameters can be adjusted to highlight desired tissues.

(\phi)

of a spin is a function of the gradient field

G(t)

:

\phi=\gamma

t
\int
0

B0+G(\tau)r(\tau)d\tau

where

\gamma

is the Gyromagnetic ratio and

r

is defined as:

r(\tau)=r0+vr\tau+

1
2

ar\tau2+\ldots

,

r0

is the initial position of the spin,

vr

is the spin velocity, and

ar

is the spin acceleration. If we only consider static spins and spins in the x-direction, we can rewrite equation for phase shift as:

\phi=\gammax0

t
\int
0

Gx(\tau)d\tau+\gammavx

t
\int
0

Gx(\tau)\taud\tau+\gamma

ax
2
t
\int
0

Gx(\tau)\tau2d\tau+\ldots

We then assume that acceleration and higher order terms are negligible to simplify the expression for phase to:

\phi=\gamma(x0M0+vxM1)

where

M0

is the zeroth moment of the x-gradient and

M1

is the first moment of the x gradient.

If we take two different acquisitions with applied magnetic gradients that are the opposite of each other (bipolar gradients), we can add the results of the two acquisitions together to calculate a change in phase that is dependent on gradient:

\Delta\phi=v(\gamma\DeltaM1)

where

\DeltaM1=2M1

.[2] [3]

The phase shift is measured and converted to a velocity according to the following equation:

v=

venc
\pi

\Delta\phi

where

venc

is the maximum velocity that can be recorded and

\Delta\phi

is the recorded phase shift.

The choice of

venc

defines range of velocities visible, known as the ‘dynamic range’.A choice of

venc

below the maximum velocity in the slice will induce aliasing in the image where a velocity just greater than

venc

will be incorrectly calculated as moving in the opposite direction. However, there is a direct trade-off between the maximum velocity that can be encoded and the signal-to-noise ratio of the velocity measurements. This can be described by:

SNRv=

\pi
\sqrt{2
} \frac SNR where

SNR

is the signal-to-noise ratio of the image (which depends on the magnetic field of the scanner, the voxel volume, and the acquisition time of the scan).

For an example, setting a ‘low’

venc

(below the maximum velocity expected in the scan) will allow for better visualization of slower velocities (better SNR), but any higher velocities will alias to an incorrect value. Setting a ‘high’

venc

(above the maximum velocity expected in the scan) will allow for the proper velocity quantification, but the larger dynamic range will obscure the smaller velocity features as well as decrease SNR. Therefore, the setting of

venc

will be application dependent and care must be exercised in the selection. In order to further allow for proper velocity quantification, especially in clinical applications where the velocity dynamic range of flow is high (e.g. blood flow velocities in vessels across the thoracoabdominal cavity), a dual-echo PC-MRI (DEPC) method with dual velocity encoding in the same repetition time has been developed.[4] The DEPC method does not only allow for proper velocity quantification, but also reduces the total acquisition time (especially when applied to 4D flow imaging) compared to a single-echo single-

venc

PC-MRI acquisition carried out at two separate

venc

values.

To allow for more flexibility in selecting

venc

, instantaneous phase (phase unwrapping) can be used to increase both dynamic range and SNR.[5]

Encoding Methods

When each dimension of velocity is calculated based on acquisitions from oppositely applied gradients, this is known as a six-point method. However, more efficient methods are also used. Two are described here:

Simple Four-point Method

Four sets of encoding gradients are used. The first is a reference and applies a negative moment in

x

,

y

, and

z

. The next applies a positive moment in

x

, and a negative moment in

y

and

z

. The third applies a positive moment in

y

, and a negative moment in

x

and

z

. And the last applies a positive moment in

z

, and a negative moment in

x

and

y

.[6] Then, the velocities can be solved based on the phase information from the corresponding phase encodes as follows:

\hat{v}x=

\phix-\phi0
\gamma\DeltaM1

\hat{v}y=

\phiy-\phi0
\gamma\DeltaM1

\hat{v}z=

\phiz-\phi0
\gamma\DeltaM1

Balanced Four-Point Method

The balanced four-point method also includes four sets of encoding gradients. The first is the same as in the simple four-point method with negative gradients applied in all directions. The second has a negative moment in

x

, and a positive moment in

y

and

z

. The third has a negative moment in

y

, and a positive moment in

x

and

z

. The last has a negative moment in

z

and a positive moment in

x

and

y

.[7] This gives us the following system of equations:

\phi2-\phi1=\phix+\phiy

\phi3-\phi1=\phix+\phiz

\phi4-\phi1=\phiy+\phiz

Then, the velocities can be calculated:

\hat{v}x=

-\phi1+\phi2+\phi3-\phi4
2\gamma\DeltaM1

\hat{v}y=

-\phi1+\phi2-\phi3+\phi4
2\gamma\DeltaM1

\hat{v}z=

-\phi1-\phi2+\phi3+\phi4
2\gamma\DeltaM1

Retrospective Cardiac and Respiratory Gating

For medical imaging, in order to get highly resolved scans in 3D space and time without motion artifacts from the heart or lungs, retrospective cardiac gating and respiratory compensation are employed. Beginning with cardiac gating, the patient's ECG signal is recorded throughout the imaging process. Similarly, the patient's respiratory patterns can be tracked throughout the scan. After the scan, the continuously collected data in k-space (temporary image space) can be assigned accordingly to match-up with the timing of the heart beat and lung motion of the patient. This means that these scans are cardiac-averaged so the measured blood velocities are an average over multiple cardiac cycles.[8]

Applications

Phase contrast MRI is one of the main techniques for magnetic resonance angiography (MRA). This is used to generate images of arteries (and less commonly veins) in order to evaluate them for stenosis (abnormal narrowing), occlusions, aneurysms (vessel wall dilatations, at risk of rupture) or other abnormalities. MRA is often used to evaluate the arteries of the neck and brain, the thoracic and abdominal aorta, the renal arteries, and the legs (the latter exam is often referred to as a "run-off").

Limitations

In particular, a few limitations of PC-MRI are of importance for the measured velocities:

Vastly undersampled Isotropic Projection Reconstruction (VIPR)

A Vastly undersampled Isotropic Projection Reconstruction (VIPR) is a radially acquired MRI sequence which results in high-resolution MRA with significantly reduced scan times, and without the need for breath-holding.[10]

Notes and References

  1. Stankovic . Zoran . Allen . Bradley D. . Garcia . Julio . Jarvis . Kelly B. . Markl . Michael . 2014 . 4D flow imaging with MRI . . 4 . 2 . 173–192 . 10.3978/j.issn.2223-3652.2014.01.02 . 24834414 . 3996243 .
  2. Elkins . C. . Alley . M.T. . 2007 . Magnetic resonance velocimetry: applications of magnetic resonance imaging in the measurement of fluid motion . Experiments in Fluids. 43 . 6 . 823 . 2007ExFl...43..823E . 10.1007/s00348-007-0383-2 . 121958168 .
  3. Taylor . Charles A. . Draney . Mary T. . 2004 . Experimental and computational methods in cardiovascular fluid mechanics . . 36 . 197–231 . 10.1146/annurev.fluid.36.050802.121944 . 2004AnRFM..36..197T .
  4. Ajala . Afis . Zhang . Jiming . Pednekar . Amol . Buko . Erick . Wang . Luning . Cheong . Benjamin . Hor . Pei-Herng . Muthupillai . Raja . 2020 . Mitral Valve Flow and Myocardial Motion Assessed with Dual-Echo Dual-Velocity Cardiac MRI . Radiology: Cardiothoracic Imaging. 3 . 2 . e190126 . 10.1148/ryct.2020190126. 33778578 . 7977974 .
  5. Salfitya . M.F. . Huntleya . J.M. . Gravesb . M.J. . Marklundc . O. . Cusackd . R. . Beauregardd . D.A. . 2006 . Extending the dynamic range of phase contrast magnetic resonance velocity imaging using advanced higher-dimensional phase unwrapping algorithms . . 3 . 8 . 415–427 . 10.1098/rsif.2005.0096 . 16849270 . 1578755 .
  6. Pelc . Norbert J. . Bernstein . Matt A. . Shimakawa . Ann . Glover . Gary H. . 1991 . Encoding strategies for three‐direction phase‐contrast MR imaging of flow . . 1 . 4 . 405–413 . 10.1002/jmri.1880010404. 1790362 . 3000911 .
  7. Pelc . Norbert J. . Bernstein . Matt A. . Shimakawa . Ann . Glover . Gary H. . 1991 . Encoding strategies for three‐direction phase‐contrast MR imaging of flow . . 1 . 4 . 405–413 . 10.1002/jmri.1880010404. 1790362 . 3000911 .
  8. Lotz . Joachim . Meier . Christian . Leppert . Andreas . Galanski . Michael . 2002 . Cardiovascular Flow Measurement with Phase-Contrast MR Imaging: Basic Facts and Implementation 1 . Radiographics. 22 . 3 . 651–671 . 10.1148/radiographics.22.3.g02ma11651. 12006694 .
  9. Petersson . Sven . Dyverfeldt . Petter . Gårdhagen . Roland . Karlsson . Matts . Ebbers . Tino . 2010 . Simulation of phase contrast MRI of turbulent flow . . 64 . 4 . 1039–1046 . 10.1002/mrm.22494 . 20574963 . free .
  10. https://books.google.com/books?id=yZ3uCQAAQBAJ&pg=PA602 Page 602