The Schlenk line (also vacuum gas manifold) is a commonly used chemistry apparatus developed by Wilhelm Schlenk.[1] It consists of a dual manifold with several ports.[2] One manifold is connected to a source of purified inert gas, while the other is connected to a vacuum pump. The inert-gas line is vented through an oil bubbler, while solvent vapors and gaseous reaction products are prevented from contaminating the vacuum pump by a liquid-nitrogen or dry-ice/acetone cold trap. Special stopcocks or Teflon taps allow vacuum or inert gas to be selected without the need for placing the sample on a separate line.[3]
Schlenk lines are useful for manipulating moisture- and air-sensitive compounds. The vacuum is used to remove air or other gasses present in closed, connected glassware to the line. It often also removes the last traces of solvent from a sample. Vacuum and gas manifolds often have many ports and lines, and with care, it is possible for several reactions or operations to be run simultaneously in inert conditions.
When the reagents are highly susceptible to oxidation, traces of oxygen may pose a problem. Then, for the removal of oxygen below the ppm level, the inert gas needs to be purified by passing it through a deoxygenation catalyst.[4] This is usually a column of copper(I) or manganese(II) oxide, which reacts with oxygen traces present in the inert gas. In other cases, a purge-cycle technique is often employed, where the closed, reaction vessel connected to the line is filled with inert gas, evacuated with the vacuum and then refilled. This process is repeated 3 or more times to make sure air is rigorously removed. Moisture can be removed by heating the reaction vessel with a heat gun. [5]
The main techniques associated with the use of a Schlenk line include:
Glassware are usually connected by tightly fitting and greased ground glass joints. Round bends of glass tubing with ground glass joints may be used to adjust the orientation of various vessels. Glassware is necessarily purged of outside air by using the purge cycling technique. The solvents and reagents that are used can use a technique called "sparging" to remove air. This is where a cannula needle, which is connected to the inert gas on the line, is inserted into the reaction vessel containing the solvent; this effectively bubbles the inert gas into the solution, which will actively push out trapped gas molecules from the solvent.
Filtration under inert conditions poses a special challenge. It is usually achieved using a "cannula filter".[3] Classically, filtration is tackled with a Schlenk filter, which consists of a sintered glass funnel fitted with joints and stopcocks. By fitting the pre-dried funnel and receiving flask to the reaction flask against a flow of nitrogen, carefully inverting the set-up and turning on the vacuum appropriately, the filtration may be accomplished with minimal exposure to air.
A glovebox is often used in conjunction with the Schlenk line for storing and reusing air- and moisture-sensitive solvents in a lab.
The main dangers associated with the use of a Schlenk line are the risks of an implosion or explosion. An implosion can occur due to the use of vacuum and flaws in the glass apparatus.
An explosion can occur due to the common use of liquid nitrogen in the cold trap, used to protect the vacuum pump from solvents. If a reasonable amount of air is allowed to enter the Schlenk line, liquid oxygen can condense into the cold trap as a pale blue liquid. An explosion may occur due to reaction of the liquid oxygen with any organic compounds also in the trap.