PET-MRI explained

Positron emission tomography–magnetic resonance imaging
Purpose:used in clinical field of oncology

Positron emission tomography–magnetic resonance imaging (PET–MRI) is a hybrid imaging technology that incorporates magnetic resonance imaging (MRI) soft tissue morphological imaging and positron emission tomography (PET) functional imaging.[1]

The combination of PET and MRI was mentioned in a 1991 Phd thesis by R. Raylman.[2] Simultaneous PET/MR detection was first demonstrated in 1997, however it took another 13 years, and new detector technologies, for clinical systems to become commercially available.[3]

Applications

Presently, the main clinical fields of PET-MRI are oncology,[4] [5] [6] cardiology,[7] neurology,[8] [9] [10] and neuroscience.[11] Research studies are actively conducted at the moment to understand benefits of the new PET-MRI diagnostic method. The technology combines the exquisite structural and functional characterization of tissue provided by MRI with the extreme sensitivity of PET imaging of metabolism and tracking of uniquely labeled cell types or cell receptors.

Manufacturers

Several companies offer clinical and pre-clinical combined PET-MR system; clinical systems are available from United Imaging, Philips, Siemens, and GE. There are varying approaches to the combination of the two technologies. Some designs are essentially separate machines, in the same room, with a bed that can transfer a patient from one scanner to another.[12] [13] Fully integrated systems are the most technically challenging to achieve, but provide greatest benefits in terms of the ability to make simultaneous, exactly aligned, acquisitions.[14] [15]

Clinical systems

The first two clinical whole body PET-MRI systems were installed by Philips at Mount Sinai Medical Centre in the United States and at Geneva University Hospital in Switzerland, in 2010. The system featured a PET and MRI scanner separated by a revolving bed.[16] [17]

Siemens was the first company to offer simultaneous PET/MR acquisitions, with the first systems installed in 2010 based on avalanche photodiode detectors.[18]

Currently Siemens and GE are the only companies to offer a fully integrated whole body and simultaneous acquisition PET-MRI system. The Siemens system (Biograph mMR) received a CE mark[19] and FDA approval[20] for customer purchase in 2011.

The GE system (SIGNA PET/MR) received its 510K & CE mark in 2014.

Preclinical systems

Currently, the combination of positron emission tomography (PET) and magnetic resonance imaging (MRI) as a hybrid imaging modality is receiving great attention not only in its emerging clinical applications but also in the preclinical field. Several designs based on several different types of PET detector technology have been developed in recent years, some of which have been used for first preclinical studies.[21] [22] [23]

Several companies offer MR-compatible preclinical PET scanner inserts for use in the bore of an existing MRI, enabling simultaneous PET/MR image acquisition.[24] [25] [26] [27]

Comparison with PET-CT

The combination of PET with X-ray computed tomography (CT) is the more established PET imaging technology. With both PET-CT and PET-MR the intended advantage is to combine functional imaging provided by PET, with structural (anatomical) information from CT or MRI. Although images from different modalities collected at different scanning sessions can be overlaid by image registration, a simultaneous acquisition offers better alignment of images and direct correlation. Combining imaging modalities in one single scanning session also has the advantage of reducing the number of appointments and therefore improving patient comfort.[28] [29]

The same clinical decisions that would influence the choice between stand-alone CT or MR imaging would also determine areas where PET-CT or PET-MR would be preferred. For example, one advantage of MRI compared to CT is its superior soft tissue contrast, while CT has the advantage of being much faster than MRI.

One clear advantage of PET-MR compared to PET-CT is the lower total ionising radiation dose obtained. For body PET-CT applications, the CT part of the examination constitutes approximately 60-80% of the radiation dose, with the remaining radiation dose originating from the PET radiopharmaceutical.[30] In contrast, no ionising radiation dose is obtained from MRI. PET-MR is therefore appealing in children, in particularly for serial follow-up examinations as used in oncology or chronic inflammatory conditions.[31]

Attenuation correction

PET-MRI systems don't offer a direct way to obtain attenuation maps, unlike stand-alone PET or PET-CT systems.[32] [33]

Stand alone PET systems' attenuation correction (AC) is based on a transmission scan (mu - map) acquired using a 68Ge (Germanium-68) rotating rod source, which directly measures photon attenuation at 511 keV.[34] PET-CT systems use a low-dose CT scan for AC. Since X-rays have a range of energies lower than 511 keV, AC values are closely approximated from Hounsfield units.[35]

There is no correlation between MR image intensity and electron intensity, therefore conversion of MR images into an attenuation map is difficult.[36] [32] [34] This is an active area of research and a range of approaches have been developed. One method uses a Dixon MRI sequence, and segments the resultant image into fat and water, with pre-set attenuation factors. Disadvantages of this method include a lack of bone attenuation, and loss of the true continuous range of attenuation factors. Comparisons with PET-CT attenuation maps for oncology purposes however have shown that this is a usable technique.[34] The Dixon method can be combined with ultrashort echo time (UTE) sequences to better identify bone and increase the possible classes of tissue for segmentation. More sequences increase MRI acquisition time, and therefore the risk of motion artefacts.[37]

In areas of the body with predictable structures (e.g. the head), segmentation (where tissue is categorised using the MRI image data), or "atlas" methods can be used. In atlas methods a standard MR image, with associated CT attenuation data, can be warped to fit the actual patient anatomy. Disadvantages of this method include difficulty with unusual anatomy, a need for a suitable library of images, and the need to account for MR coil attenuation.[34] Synthetic, or Substitute CT (sCT) methods to generate CT like data from MRI are also of interest for radiotherapy planning, and have been primarily investigated for sites in the head. While some of these use an atlas technique, many take a voxel approach where actual voxel intensities (contrast data) are used in combination with machine learning (trained on MR/CT data) to assign electron density values.[34] [38] [39]

In many of the above methods, MRI artifacts (e.g. from physiological motion) can affect attenuation correction accuracy.[34] [40]

See also

Notes and References

  1. Antoch. Gerald. Bockisch. Andreas. Combined PET/MRI: a new dimension in whole-body oncology imaging?. European Journal of Nuclear Medicine and Molecular Imaging. 36. S1. 2008. 113–120. 1619-7070. 10.1007/s00259-008-0951-6. 19104802. 8153201.
  2. Web site: Reduction of positron range effects by the application of a magnetic field: For use with positron emission tomography . .
  3. Book: Luna . Antonio . Vilanova . Joan C. . Hygino da Cruz . L. Celso Jr. . Rossi . Santiago E. . Functional Imaging in Oncology: Biophysical Basis and Technical Approaches . 2013 . Springer Science & Business Media . 9783642404122 . 421 .
  4. Buchbender C . 4 . Heusner TA . Lauenstein TC. Bockisch A . Antoch G . Oncologic PET/MRI, part 1: tumors of the brain, head and neck, chest, abdomen, and pelvis . Journal of Nuclear Medicine . 53 . 6 . 928–38 . June 2012 . 22582048 . 10.2967/jnumed.112.105338. free .
  5. Buchbender C . 4 . Heusner TA . Lauenstein TC. Bockisch A . Antoch G . Oncologic PET/MRI, part 2: bone tumors, soft-tissue tumors, melanoma, and lymphoma . Journal of Nuclear Medicine . 53 . 8 . 1244–52 . August 2012 . 22782313 . 10.2967/jnumed.112.109306. free .
  6. Martinez-Möller A . Eiber M . Nekolla SG . etal . Workflow and scan protocol considerations for integrated whole-body PET/MRI in oncology . Journal of Nuclear Medicine . 53 . 9 . 1415–26 . September 2012 . 22879079 . 10.2967/jnumed.112.109348. free .
  7. Rischpler C . Nekolla SG . Dregely I. Schwaiger M . Hybrid PET/MR imaging of the heart: potential, initial experiences, and future prospects . Journal of Nuclear Medicine . 54 . 3 . 402–15 . March 2013 . 23404088 . 10.2967/jnumed.112.105353. free .
  8. Web site: Imaging of traumatic brain injury patients swifter and safer with new technology at NIH . 18 September 2015 .
  9. Dimou E . Booij J . Rodrigues M . etal . Amyloid PET and MRI in Alzheimer's disease and mild cognitive impairment . Current Alzheimer Research . 6 . 3 . 312–9 . June 2009 . 19519314 . 10.2174/156720509788486563.
  10. Bremner JD . Vythilingam M . Vermetten E . etal . MRI and PET study of deficits in hippocampal structure and function in women with childhood sexual abuse and posttraumatic stress disorder . The American Journal of Psychiatry . 160 . 5 . 924–32 . May 2003 . 12727697 . 10.1176/appi.ajp.160.5.924.
  11. Cho . Zang Hee . Son . Young Don . Choi . Eun Jung . Kim . Hang Keun . Kim . Jeong Hee . Lee . Sang Yoon . Ogawa . Seiji . Kim . Young Bo . In-vivo human brain molecular imaging with a brain-dedicated PET/MRI system . Magnetic Resonance Materials in Physics, Biology and Medicine . 3 August 2012 . 26 . 1 . 71–79 . 10.1007/s10334-012-0329-4. 22864642 . 10721235 .
  12. Torigian . Drew A. . Zaidi . Habib . Kwee . Thomas C. . Saboury . Babak . Udupa . Jayaram K. . Cho . Zang-Hee . Alavi . Abass . PET/MR Imaging: Technical Aspects and Potential Clinical Applications . Radiology . April 2013 . 267 . 1 . 26–44 . 10.1148/radiol.13121038 . 23525716.
  13. Web site: The Past, Present and Future of PET/MRI Scanners . Imaging Technology News . 15 January 2019 . 5 May 2017.
  14. Jadvar . Hossein . Colletti . Patrick M. . Competitive advantage of PET/MRI . European Journal of Radiology . January 2014 . 83 . 1 . 84–94 . 10.1016/j.ejrad.2013.05.028 . 23791129 . 3800216.
  15. Mannheim . Julia G. . Schmid . Andreas M. . Schwenck . Johannes . Katiyar . Prateek . Herfert . Kristina . Pichler . Bernd J. . Disselhorst . Jonathan A. . PET/MRI Hybrid Systems . Seminars in Nuclear Medicine . July 2018 . 48 . 4 . 332–347 . 10.1053/j.semnuclmed.2018.02.011. 29852943 . 46924919 .
  16. Web site: Wood . Harry . PET-MRI scanner opens new frontier in medical imaging . Medical Technology Business Europe . 15 January 2019 . 28 May 2010.
  17. Muzic . Raymond F. . DiFilippo . Frank P. . Positron Emission Tomography-Magnetic Resonance Imaging: Technical Review . Seminars in Roentgenology . July 2014 . 49 . 3 . 242–254 . 10.1053/j.ro.2014.10.001 . 25497909 . 4451572.
  18. Book: Zaidi . Habib . PET/MRI: Advances in Instrumentation and Quantitative Procedures, An Issue of PET Clinics . 2016 . Elsevier Health Sciences . 9780323417686 .
  19. Web site: Siemens receives CE mark for whole-body molecular MR system . Healthcare Sector, Siemens AG . 2011-06-01 . 2014-01-05.
  20. Web site: FDA clears new system to perform simultaneous PET, MRI scans . U.S. Food and Drug Administration . 2011-06-10 . 2014-01-04.
  21. 10.1053/j.semnuclmed.2012.08.004 . Applications for Preclinical PET/MRI . 2013 . Judenhofer . Martin S. . Cherry . Simon R. . Seminars in Nuclear Medicine . 43 . 19–29 . 23178086 . 1.
  22. Book: 10.1109/NSSMIC.2011.6152496 . 2011 . Schulz . Volkmar . Weissler . Bjoern . Gebhardt . Pierre . Solf . Torsten . Lerche . Christoph . Fischer . Peter . Ritzert . Michael . Piemonte . Claudio . Goldschmidt . Benjamin . Vandenberghe . Stefaan . Salomon . Andre . Schaeffter . Tobias . Marsden . Paul . 2011 IEEE Nuclear Science Symposium Conference Record . SiPM based preclinical PET/MR insert for a human 3T MR: First imaging experiments . 4467–4469. 978-1-4673-0120-6 . 27832030 .
  23. 10.1016/j.nima.2013.08.077 . 25843999 . 4376059 . PET/MRI insert using digital SiPMs: Investigation of MR-compatibility . 2013 . Wehner . Jakob . Weissler. Bjoern . Dueppenbecker. Peter . Gebhardt. Pierre . Schug. David . Ruetten. Walter . Kiessling. Fabian . Schulz. Volkmar . Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 2014NIMPA.734..116W . 734 . Pt B . 116–121.
  24. Omidvari . Negar . Cabello . Jorge . Topping . Geoffrey . Schneider . Florian Roland . Paul . Stephan . Schwaiger . Markus . Ziegler . Sibylle I . PET performance evaluation of MADPET4: a small animal PET insert for a 7-Tesla MRI scanner . Physics in Medicine and Biology . 62 . 22 . 8671–8692 . 4 October 2017 . 10.1088/1361-6560/aa910d. 28976912 . 4352807 .
  25. Wehner . J . Weissler . B . Dueppenbecker . P M . Gebhardt . P . Goldschmidt . B . Schug . D . Kiessling . F . Schulz . V . MR-compatibility assessment of the first preclinical PET-MRI insert equipped with digital silicon photomultipliers . Physics in Medicine and Biology . 21 March 2015 . 60 . 6 . 2231–2255 . 10.1088/0031-9155/60/6/2231. 25684065 . 2015PMB....60.2231W . free .
  26. Goldenberg . Joshua M. . Cárdenas-Rodríguez . Julio . Pagel . Mark D. . Preliminary Results that Assess Metformin Treatment in a Preclinical Model of Pancreatic Cancer Using Simultaneous [18F]FDG PET and acidoCEST MRI . Molecular Imaging and Biology . 26 January 2018 . 20 . 4 . 575–583 . 10.1007/s11307-018-1164-4 . 29374343 . 6043393.
  27. 10.2967/jnumed.112.119065 . 23990683 . Performance Evaluation of the Small-Animal nanoScan PET/MRI System . 2013 . Nagy . Kálmán . Tóth . Miklós . Major . Péter . Patay . Győző . Egri . G. . Häggkvist . Jenny . Varrone. Andrea . Farde . Lars . Halldin . Christer . Gulyás . Balázs . Journal of Nuclear Medicine . 54 . 10 . 1825–1832. free .
  28. Web site: Kaplan . Deborah Abrams . PET/MRI: Reflections Two Years After FDA Approval . Diagnostic Imaging . 15 January 2019 . 12 June 2013.
  29. 2762705 . 18396179 . 10.1053/j.semnuclmed.2008.02.001 . 38 . 3 . Positron emission tomography/magnetic resonance imaging: the next generation of multimodality imaging? . 2008 . Semin Nucl Med . 199–208 . Pichler BJ, Wehrl HF, Kolb A, Judenhofer MS . Bernd Pichler.
  30. Martí-Climent. Josep M.. Prieto. Elena. Morán. Verónica. Sancho. Lidia. Rodríguez-Fraile. Macarena. Arbizu. Javier. García-Velloso. María J.. Richter. José A.. December 2017. Effective dose estimation for oncological and neurological PET/CT procedures. EJNMMI Research. 7. 1. 37. 10.1186/s13550-017-0272-5. 28439843. 5403773. 2191-219X . free .
  31. Ehman. Eric C.. Johnson. Geoffrey B.. Villanueva-Meyer. Javier E.. Cha. Soonmee. Leynes. Andrew Palmera. Larson. Peder Eric Zufall. Hope. Thomas A.. November 2017. PET/MRI: Where might it replace PET/CT?. Journal of Magnetic Resonance Imaging. 46. 5. 1247–1262. 10.1002/jmri.25711. 5623147. 28370695.
  32. Keereman. Vincent. Mollet. Pieter. Berker. Yannick. Schulz. Volkmar. Vandenberghe. Stefaan. 2013-02-01. Challenges and current methods for attenuation correction in PET/MR. Magnetic Resonance Materials in Physics, Biology and Medicine. 26. 1. 81–98. 10.1007/s10334-012-0334-7. 22875599. 22198626. 0968-5243.
  33. van Dalen. Jorn A.. Visser. Eric P.. Vogel. Wouter V.. Corstens. Frans H. M.. Oyen. Wim J. G.. 2007-03-01. Impact of Ge-68/Ga-68-based versus CT-based attenuation correction on PET. Medical Physics. 34. 3. 889–897. 10.1118/1.2437283. 17441234. 2473-4209. 2007MedPh..34..889V.
  34. Wagenknecht. Gudrun. Kaiser. Hans-Jürgen. Mottaghy. Felix M.. Herzog. Hans. 2013-02-01. MRI for attenuation correction in PET: methods and challenges. Magnetic Resonance Materials in Physics, Biology and Medicine. 26. 1. 99–113. 10.1007/s10334-012-0353-4. 23179594. free. 3572388. 0968-5243.
  35. Bai. Chuanyong. Shao. Ling. Silva. A. J. Da. Zhao. Zuo. October 2003. A generalized model for the conversion from CT numbers to linear attenuation coefficients. IEEE Transactions on Nuclear Science. 50. 5. 1510–1515. 10.1109/tns.2003.817281. 0018-9499. 2003ITNS...50.1510B.
  36. Hofmann. Matthias. Pichler. Bernd. Schölkopf. Bernhard. Beyer. Thomas. 2009-03-01. Towards quantitative PET/MRI: a review of MR-based attenuation correction techniques. European Journal of Nuclear Medicine and Molecular Imaging. 36. 1. 93–104. 10.1007/s00259-008-1007-7. 19104810. 1619-7070. free. 11858/00-001M-0000-0013-C595-F. free.
  37. Vandenberghe . Stefaan . Marsden . Paul K . PET-MRI: a review of challenges and solutions in the development of integrated multimodality imaging . Physics in Medicine and Biology . 21 February 2015 . 60 . 4 . R115–R154 . 10.1088/0031-9155/60/4/R115 . 25650582. 1510.04875 . 2015PMB....60R.115V . free.
  38. Edmund . Jens M. . Nyholm . Tufve . A review of substitute CT generation for MRI-only radiation therapy . Radiation Oncology . 26 January 2017 . 12 . 1 . 28 . 10.1186/s13014-016-0747-y . 28126030 . 5270229 . free .
  39. Larsson . Anne . Johansson . Adam . Axelsson . Jan . Nyholm . Tufve . Asklund . Thomas . Riklund . Katrine . Karlsson . Mikael . Evaluation of an attenuation correction method for PET/MR imaging of the head based on substitute CT images . Magnetic Resonance Materials in Physics, Biology and Medicine . 7 September 2012 . 26 . 1 . 127–136 . 10.1007/s10334-012-0339-2 . 22955943. 7334804 .
  40. Hofmann . Matthias . Pichler . Bernd . Schölkopf . Bernhard . Beyer . Thomas . Towards quantitative PET/MRI: a review of MR-based attenuation correction techniques . European Journal of Nuclear Medicine and Molecular Imaging . 23 December 2008 . 36 . S1 . 93–104 . 10.1007/s00259-008-1007-7 . 19104810. free. 11858/00-001M-0000-0013-C595-F . free .