A monitor unit (MU) is a measure of machine output from a clinical accelerator for radiation therapy such as a linear accelerator or an orthovoltage unit. Monitor units are measured by monitor chambers, which are ionization chambers that measure the dose delivered by a beam and are built into the treatment head of radiotherapy linear accelerators.[1]
Linear accelerators are calibrated to give a particular absorbed dose under particular conditions, although the definition and measurement configuration may vary among medical clinics.[2] [3]
The most common definitions are:[4]
Some linear accelerators are calibrated using source-to-axis distance (SAD) instead of source-to-surface distance (SSD), and calibration (monitor unit definition) may vary depending on hospital custom.
Early radiotherapy was performed using "constant SSD" treatments, and so the definition of monitor unit was adopted to reflect this calibration geometry.
Modern radiotherapy is performed using isocentric treatment plans, so newer definitions of the monitor unit are based on geometry at the isocenter based on the source-to-axis distance (SAD).
Nearly 60% of the reported errors involved a lack of an appropriate independent secondary check of the treatment plan or dose calculation [5]
With the development and technological advances, radiotherapy requires that high doses of radiation are delivered to the tumor with increasing precision. According to the recommendations of the International Commission on Radiation Units and Measurements (ICRU) in Publication 24,[6] the delivered dose should not deviate by more than ± 5% of the prescribed dose. More recently, the new ICRU recommendations in Publication 62[7]
Commercially available computerized treatment planning systems are often used in radiotherapy services to perform monitoring unit (MU) calculations to deliver the prescribed dose to the patient. As only a part of the total dose uncertainty originates from the calculation process in treatment planning, the tolerance for accuracy of planning systems has to be smaller.[8] [9]
Publications on quality assurance in radiotherapy have recommended routine checks of MU calculations through independent manual calculation. This type of verification can also increase confidence in the accuracy of the algorithm and in the data integrity of the beams used, in addition to providing an indication of the limitations of the application of conventional dose calculation algorithms used by planning systems.[10] Table 1 lists MU calculation software manufacturers.
LAP Laser | RadCalc | Modified Clarkson | IMRTVMAT TomoTherapy CyberKnife Halcyon | https://www.lap-laser.com/ | |
Varian Medical Systems | Mobius | Collapsed Cone Convolution/Superposition | IMRTVMAT TomoTherapy CyberKnife Halcyon | https://www.varian.com/ | |
Standard Imaging | IMSure | Three Source Model | IMRTVMAT | https://www.standardimaging.com/ | |
PTW Freiburg | Diamond | Modified Clarkson | IMRTVMAT | https://www.ptwdosimetry.com/en/ | |
Sun Nuclear | DoseCHECK | Collapsed Cone Convolution/Superposition | IMRTVMAT TomoTherapy Halcyon | https://www.sunnuclear.com/ | |
RT Medical Systems | RT Connect | Modified Clarkson | 3D, 2DIMRT VMAT TomoTherapy CyberKnife Halcyon | http://rtmedical.com.br/ | |
Math Resolutions LLC | DosimetryCheck | Modified Clarkson | IMRTVMAT | http://www.mathresolutions.com/ | |
MUCheck | Oncology Data Systems | Modified Clarkson | IMRTVMAT TomoTherapy CyberKnife | https://mucheck.com/odshome/ |